Print
Bimedica.
Ref. Package Packaging
101020 1000 ml 12 IU
101021 5000 ml 4 IU
101022 Reactive Strips 1 IU

Lysonox®PAA High-level disinfectant and cold chemical sterilisation

High-level sporicidal instrument disinfectant

Medical Device Class IIb - According to Regulation (EU) MDR 2017/745

Indications for use:

Step 1: activation of the solution

Pour the activator into the generator (1L or 5L) to activate the disinfectant solution. Shake lightly to mix the components and wait 30 minutes for activation.

Step 2: use of the activated solution.

In a container, pour the activated solution and soak the cleaned material for 15 minutes to disinfect it.

Step 3: Completion of disinfection.

Rinse the disinfectant with sterile water and dry, maintaining sterile conditions to ensure complete disinfection of the material.

Features and Benefits:
  • Requires activation
  • Stability of activated solution: 18 days.
  • Wide spectrum of activity.
  • Contact time between 5 and 15 minutes.
  • Compatible with stainless steel, aluminium, polyethylene, PVC, polystyrene, Teflon and polycarbonates.
  • Not recommended for materials such as copper or zinc.
  • Does not require dilution.
  • Compatible with test strips (ref 101022) to check if the solution is activated.

Composition

Peracetic acid and Tea Tree Oil

Standards

EN 14561, EN 14562, EN 14476, EN 13727. EN 13624, EN 14348, EN 17111, EN 17126